Menu
Your Cart

Agilent Modules

Brand: HP Product Code: 10596
Pressure Range: 0-40 MPa (0-400 Bar, 0-5880 PSI)
Solvent channels: 4 Channels
Flow Rate Range: 0.1-5.0 ml/min (0.0001-0.005 L/min, 100-5000 uL/min)
Resolution: 0.1 ml/min (0.0001 L/min, 100 uL/min)
Wetted Materials: Stainless Steal, gold, sapphire, ceramic, ruby, PTFE, tantalum
Brand: HP Product Code: 10782
Hydraulic system: Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston desig
Pressure Operating range: 0 to 400 bar
Solvent Channels: 4
Flow range: 0.1 – 5.0 ml/min
Settable flow range Setpoints: 0.001 – 5 ml/min, in 0.001 ml/min increments
Brand: Agilent Product Code: 11001
Detector type: DAD
Standard: 13 ul volume, 10 mm cell path length and 120 bar (1760 psi) pressure maximum
Analog outputs Recorder/integrator: 100 mV or 1 V, output range0.001 – 2 AU, two outputs
Wavelength range: 190 – 950 nm, Linear absorbance range > 2 AU (upper limit)
Chromatography Type: HPLC
Brand: HP Product Code: 10998
Carry Over Typically <0.004 %
Capacity: 100 Micro vial tray
Injection range: 0.1 – 100 uL in 0.1 uL increments
Sample viscosity range: 0.2 – 5 cp
Brand: Shimadzu Product Code: 29257
Measurement Method: Light Scattering, Detection: Photomultiplier Tube
Temperature Setting Range: ambient to 80C
Operating Temperature Range: 5 - 40C, Humidity Range: <80% (5 - 31C), <50% (31 - 40C)
Gas Flow Rate: max 3.0 L/min, Gas Pressure: max 450 kPa
Flow Rate (Mobile Phase): 0.2mL/min - 2.5mL/min
Brand: Sedere Product Code: 14025
Detection: Photomultiplier (PMT)
Light Source: Blue LED Elapsed Time Counter
Temperature Range: ambient to 100C
Eluent Flow Rate: 5uL/min to 5mL/min
Chromatography Type: HPLC
Brand: Polymer Laboratories Product Code: 29238
Measurement Method: light scattering, Detection: photomultiplier
Temperature Range: evaporator 10-80C, nebulizer ambient to 90C
Nebulizer Gas: typical inlet pressure 60-100psi, maximum inlet pressure 100psi, gas flow rate 0-3.25SLM
Eluent Flow Rate: 0-5ml/min
Brand: Agilent Product Code: 13435
Hydraulic system: Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston desig
Pressure Operating range: 0 to 400 bar
Solvent Channels: 4
Flow range: 0.1 - 5.0 ml/min
Setable flow range Setpoints: 0.001 - 5 ml/min, in 0.001 ml/min increments
Brand: Agilent Product Code: 13585
Hydraulic system: Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston desig
Pressure Operating range: 0 to 400 bar
Solvent Channels: 4
Flow range: 0.1 - 5.0 ml/min
Setable flow range Setpoints: 0.001 - 5 ml/min, in 0.001 ml/min increments
Brand: Hewlett Packard Product Code: 12350
Hydraulic system: Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston desig
Pressure Operating range: 0 to 400 bar
Solvent Channels: 4
Flow range: 0.1 - 5.0 ml/min
Setable flow range Setpoints: 0.001 - 5 ml/min, in 0.001 ml/min increments
Brand: Agilent Product Code: 13819
Sample capacity: 2 x MTP well plates + 10 x 2ml, 100 x 2ml in one tray, 40 x 2ml in half tray
Injection Range: 0.01 to 8ul in 0.01ul increments with small loop capillary, 0.01 to 40ul in 0.01ul increments with extended loo
Precision: Typically < 0.5% RSD of peak areas from 5 to 40ul, Typically < 1% RSD from 1 to 5 ul, Typically < 3% RSD fro
Sample Viscosity Range: 0.2 to 5 cp
Injection cycle time: Typically < 30 sec using standard conditions (4ul/min draw speed, 10ul/min eject speed, 0.1ul injection
Brand: Agilent Product Code: 13748
Sample capacity: 2 x well-plates (MTP) + 10 x 2 ml vials, 100 x 2 ml in one tray, 40 x 2 ml in half tray
Injection range: 0.1 - 100 ul in 0.1 ul increments, Up to 1500 ul with multiple draw (hardware modification required)
Precision: Typically < 0.5 % RSD of peak areas from 5 - 100 ul, Typically < 1 % RSD from 1 - 5 ul
Sample viscosity range: 0.2 - 5 cp
Injection cycle time: Typically < 30 s using following standard conditions, Default draw speed: 200 ul/min, Default eject spe
Brand: Agilent Product Code: 13696
Number of Samples: up to 126 vials (16x100mm, 16x48mm)
Operating Altitude: up to 2000 m (6500 ft)
Humidity: < 95 %, @ 25 - 40 C (77 - 104 F)
Ambient operating temperature: 4 - 45 C (41 - 113 F)
Brand: Agilent Product Code: 13751
Hydraulic system: Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston desig
Pressure Operating range: 0 to 400 bar
Solvent Channels: 4
Flow range: 0.1 - 5.0 ml/min
Setable flow range Setpoints: 0.001 - 5 ml/min, in 0.001 ml/min increments
Brand: Agilent Product Code: 13393
Detection type: 1024-element photodiode array, Light source: Deuterium and tungsten lamps.
Wavelength range: 190-950 nm
Short term noise: +/- 1*10-5 AU at 254 and 750 nm, Drift: 2*10-3 AU/hr at 254 nm, Linear absorbance range > 2 AU (upper limit
Wavelength accuracy: +/- 1 nm, Wavelength bunching 1*400 nm
Slit width: 1, 2, 4, 8, 16 nm, Diode width < 1 nm
Brand: Agilent Product Code: 13437
Detection type: 1024-element photodiode array, Light source: Deuterium and tungsten lamps
Wavelength range: 190 - 950 nm
Short term noise: +/- 10-5 AU at 254 and 750 nm, Drift: 2 10-3 AU/hr at 254 nm, Linear absorbance range > 2 AU (upper l
Wavelength accuracy: +/- 1 nm, Wavelength bunching 1 - 400 nm
Slit width: 1, 2, 4, 8, 16 nm, Diode width < 1 nm
Brand: Agilent Product Code: 10796
Hydraulic system : Two dual piston in series, with proprietary servo-controlled variable stroke drive, floating piston, active i
Pressure range : 20 to 400 bar (5880 psi) system pressure
Solvent channels : 4
Settable column flow range : 0.01 - 4 ul/min, 1 - 2500 ul/min (with the electronic flow control bypassed)
Composition precision : less than 0.2 percent SD, at 500 nl/min (default settings), Minimum primary flow/pump channel is 5 ul/mi
Brand: Agilent Product Code: 10795
Hydraulic system : Two dual piston in series, with proprietary servo-controlled variable stroke drive, floating piston, active i
Pressure range : 20 to 400 bar (5880 psi) system pressure
Solvent channels : 4
Settable column flow range : 0.01 - 4 ul/min, 1 - 2500 ul/min (with the electronic flow control bypassed)
Composition precision : less than 0.2 percent SD, at 500 nl/min (default settings), Minimum primary flow/pump channel is 5 ul/mi
Brand: Agilent Product Code: 12998
Two independent pump channels, (two solvents per pump channel), electronic flow control for flow rates up to 4 uL/min: settable
Pressure range 20 - 400 bar (5880 psi) system pressure
Optimum composition range - 1 - 99 % or 5 uL/min per channel (primary flow), whatever is greater
Communications Controller-area network (CAN), RS-232C, APG Remote: ready, start, stop and shut-down signals
Dimensions (height-width-depth): 180 x 345 x 435 mm (7 x 13.5 x 17 inches)
Product Code: 13698
Detector type: VWD
Standard: 14-ul volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum
Recorder/integrator: 100 mV or 1 V, output range 0.001 - 2 AU, one output
Wavelength range: 190 - 600 nm
Chromatography Type: HPLC
Brand: Agilent Product Code: 11003
Detector type: VWD
Standard: 14-ul volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum
Recorder/integrator: 100 mV or 1 V, output range 0.001 - 2 AU, one output
Wavelength range 190-600 nm
Chromatography Type: HPLC
Brand: Agilent Product Code: 13436
Sample capacity: 2 well-plates (MTP) + 10 2 ml vials, 100 x 2 ml in one tray, 40 x 2 ml in half tray
Injection range: 0.1 - 100 ul in 0.1 µl increments, Up to 1500 ul with multiple draw (hardware modification required); Precision
Injection cycle time: Typically < 30 s using following standard conditions, Default draw speed: 200 ul/min, Default eject spe
Carry-over: Typically < 0.01 % using the following conditions, Column: 125 x 4 mm Hypersil ODS, 5 um, Mobile phase: Water/Ace
Communications: Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial
Showing 1 to 25 of 29 (2 Pages)