- Stock: In Stock
- Brand: Thermo
- Product Code: 49791
- Condition: Pre-Owned
Here is a Thermo Vanquish VWD Variable Wavelength Detector F VF-F40-A
SOLD AS IS / NO WARRANTY / NO RETURNS OR REFUNDS
Includes Flow Cell 6077.0360
2.5uL 7mm 120bar
Made in 2017
Condition:
This Thermo Vanquish VWD Variable Wavelength Detector is in great condition
Specifications:
Conductivity Wetted Parts: (Flow Cells) Standard and semi-micro flow cell: Fused Silica, PEEK, PTFE, Stainless Steel Bio, semipreparative and UV monitor flow cells: Fused Silica, PEEK
Description: Vanquish VWD-F, up to four channels available, 6 different flow cells available
Data Collection Rate: 250 Hz Max.
Detection Mode: UV-vis Detection (Up to 4 Channels)
Detector Type: Variable Wavelength (VWD)
Type: Variable Wavelength Detector
Drift: <0.1 mAU/h at 254 nm and dry standard flow cell
For Use With (Equipment): Vanquish UHPLC Systems
GLP Features: Predictive Performance functions: lamp age and ignitions, lamp intensity degradation, leak detection, service monitoring period. All system parameters logged in the Chromeleon CDS Audit Trail.
Lamp: Deuterium lamp, tungsten lamp
Linearity: <5% (typical <3%) at 2.5 AU caffeine, wavelength: 272 nm based on ASTM
Noise
One Channel: <±2.5 µAU at 254 nm, time constant 2 s, at 1 mL/min with water
Multiple Channels: <±10 µAU (typical <±7 µAU) at 254 nm and 280 nm, time constant 2 s, at 1 mL/min with water
Optical Bandwidth: 6 nm at 254 nm
PC Connection: USB 2.0; 3-port HUB to connect further Vanquish modules
Safety Features: Power-up diagnostics of optics, cooling fans, motors, and electronics; Leak detection and safe leak handling
Wavelength Accuracy: ±1 nm; Wavelength repeatability: ±0.1 nm
Wavelength Range: 190–900 nm with deuterium and tungsten lamp
Length (Metric): 620 mm
Width (Metric): 420 mm
Height (Metric): 160 mm
Weight (Metric): 16 kg
Electrical Requirements: 100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Please do not hesitate to contact us if you have any questions.