This Harvard BTX ECM 2001 Electro Cell Manipulator is in good working order.

Please see the exact video and photos below.

Testing we performed:
The Harvard BTX ECM 2001 Electro Cell Manipulator passed all tests and was found to be fully functional:
Tested power on: PASSED
Tested automatic and manual start buttons: PASSED
Tested potentiometers: PASSED
Tested connection to oscilloscope: PASSED
Tested controls and output: PASSED

Using an oscilloscope we tested the output vs the set points and found the Harvard BTX ECM 2001 Electro Cell Manipulator to be operating as designed.
Due to our testing you can purchase with confidence knowing this Harvard BTX ECM 2001 Electro Cell Manipulator will operate as shown and is ready to go.

System consists of:
Harvard BTX ECM 2001 Electro Cell Manipulator
Tektronix TDS 1002B Oscilloscope
BTX Probe Interface
BTX High Voltage Probe
Communication and power cables.

Configuration and Setup:
Standard Capabilities
Operating Temperature: 10 to 40C
Storage Temperature: 10 to 50C
Humidity: 90% relative humidity
Altitude: 2,000m (operating)
Pollution Degree: 2 (not to be operated in conductive pollutants atmosphere)
Power Requirements: 100 to 120VAC, 50/60 Hz
Input Power: 85W maximum (idle), 1.6kW maximum (peak)
Fusing: T, 8A, 250V 5x20mm
Dimensions (LxWxH) 43.2 x 39.4 x 28cm (17 x 15.5 x 11in), Weight 22kg (47 lbs)

Oscilloscope Power Requirements:
Input: 100-240 (VRMS)
Wattage: less than 30W

Electrical Characteristics
Alignment
Frequency: 1MHz, fixed
Voltage: 0 to 75 VRMS
Duration: 0 to 99 sec

Post-Fusion AC
Frequency: 1MHz, fixed
Voltage: (VRMS) 1/10 of the alignment AC amplitude, attenuates with time
Duration: 0 to 9sec

Electroporation
High Voltage Mode (HV Mode): Voltage 10 to 3000V peak, Pulse Length 1 to 99usec
Low Voltage Mode (LV Mode): Voltage 10 to 500V peak, Pulse Length 0.01 to 0.99msec
Number of Repeats: 0 to 9

The Harvard BTX ECM 2001 Electro Cell Manipulator is a multi functional electro cell fusion and electroporation generator.
Embryo and cell fusion is facilitated by the combination of AC and DC wave pulses. Fusion is achieved by the generation of a proprietary AC wave form for benign

Video Description:
The video starts with a wide front view of the Harvard BTX ECM 2001 Electro Cell Manipulator system.
Then we walk around of the setup, followed by a test of the controls on the ECM.
The video ends with an auto test performed by the Harvard BTX ECM 2001 Electro Cell Manipulator and recorded on the oscilloscope.

Documentation:
Datasheet: N/A

Brochure: N/A

Manual: https://drive.google.com/file/d/1KzJtxz8N-LkeTYv4Mu-VL1OSQ6qZ2v2d/view?usp=sharing

Please do not hesitate to contact us if you have any questions.
 

 

 


Harvard BTX ECM 2001 Electro Cell Manipulator Electroporator w/ Oscilloscope SOLDOUT

Manufacturer Description and Specification:

Manufacturer Documentation:

Boston Industries, Inc. © 2021