Categories

Gilson GX-281 Liquid Handler w/ 333,334 Pumps, UV/VIS-155 Detector HPLC SOLDOUT

Brand:Gilson
Product Code: 18070
Availability: SOLD
 
Qty:        

Ask a question about this product


Gilson GX-281 Liquid Handler w/ 333,334 Pumps, UV/VIS-155 Detector HPLC SOLDOUT


Send

Gilson GX-281 Liquid Handler w/ 333,334 Pumps, UV/VIS-155 Detector HPLC SOLDOUT | Priced $0.00 | bostonind.com (18070)

Here is a Gilson GX-281 Liquid Handler w/ 333,334 Pumps, UV/VIS-155 Detector HPLC

New system sold for over $100K

Predecessor to:
Current model.

Condition:
This Gilson GX-281 Liquid Handler is in good working order.

Please see the exact video and photos below.

Testing we performed:
The Gilson GX-281 Liquid Handler passed all tests and was found to be fully functional:
Tested liquid handler XYZ axis: PASSED
Tested connection with GSIOC utility: PASSED
Tested needle up/down and aspirate/dispense: PASSED
Tested UV detector using first distilled water, followed by an Di/Nai solution, values were within acceptable range: PASSED
Pump heads disassembled and cleaned via ultrasonic.
Tested 333-334 pumps manual functions (prime/flow/outlets): PASSED
Tested valves opened and closed as intended when prompted: PASSED
Tested prime function: PASSED
Tested flow and pressure functions under variable settings (max/ min flow): PASSED
Tested under pressure for leaks: PASSED

The testing performed on this Gilson GX-281 Liquid Handler and accessories confirm proper operation and functionality.
The Gilson GX-281 Liquid Handler will work properly when paired with software and accessories.

System consists of:
Gilson GX-281 Liquid Handling system
333 Prep Scale Pump w/ H3M headers
334 Prep Scale Pump w/ H3M headers
UV/VIS-155 Detector w/ 0.05mm prep scale flow cell
GX Direct Injection Module
GX Prep Solvent System Valve
Dell Latitude i5 running GSUTIL32 software
Communication and power cables.

Racks are available in several configurations please inquire.

Important: YOU WILL NEED TO PURCHASE OR HAVE GILSON TRILUTION SOFTWARE TO OPERATE THIS SYSTEM

Configuration and Setup:
Sampler Type: X/Y/Z with stationary rack design
Probe Positioning Performance: Accuracy: +/- 0.05 mm in X/Y/Z, Repeatability: +/- 0.025 mm in X/Y/Z
GX Solvent System: 50psi back pressure, flow rate preparative 1 uL/min up to 50 mL/min
UV/VIS 155 Detector: 0.05mm preparative flow cell
Software: commands can be issued via GSIOC or RS-232

UV/VIS 155 detector, Flow cell: 0.05mm Preparative
333/334 pump heads: H3M
The 333/334 pumps provide a preparative pumping solution to accommodate columns ranging from 10mm to 50mm ID in high pressure normal and reverse phases at flow rate from 1 mL/min to 200 mL/min and pressures up to 3040 psi (210 bar).
The Direct injection module comes equipped with the stainless steel valve head and a 25ml sample loop
The Z injection valve comes equipped with a 10ml sample loop

Video Description:
In this video we do a full inspection of the Gilson GX-281 Liquid Handler system paying special attention to any connections or serial tags.
The internal components of the pumps and the UV detector are shown in brief detail.
The UV lamp is seen lighting and the X,Y,Z axes are all shown traveling to their respective limits and then re-homing.
Once all relevant portions of the system have been seen we return to a full view of the Gilson GX-281 Liquid Handler system and then the item tag, after which, the video ends.

Documentation:
Datasheet: https://drive.google.com/file/d/1H-shuiF2rpDZFP0me6kafIo0pO5g8OXi/view?usp=sharing

Brochure: N/A

Manual: https://drive.google.com/file/d/1wr--xOgsxe6udxzFjHpm0sMJ-_-AaWtM/view?usp=sharing

This will ship via freight. Please contact us with your zip code to obtain shipping quote.

Please do not hesitate to contact us if you have any questions.
 

 

 


Gilson GX-281 Liquid Handler w/ 333,334 Pumps, UV/VIS-155 Detector HPLC SOLDOUT

Manufacturer Description and Specification:

Manufacturer Documentation:

Boston Industries, Inc. © 2020