Here is a BD FACSCalibur Flow Cytometer w/ HTS, FACSFlow

New costs over $100K

Predecessor to:
Current model.

Condition:
This BD FACSCalibur Flow Cytometer is in good working order.

Please see the exact video and photos below.

Testing we performed:
The BD FACSCalibur Flow Cytometer passed all tests and was found to be fully functional:
Performed cleaning of Fluidics system
Tested operation of Fluidics: PASSED
Tested operation of Optics: PASSED
Tested communications with Apple computer: PASSED
Tested communication and all functions with the HTS system: PASSED
Tested operation of the FACSFlow system: PASSED
Passed Cytek Calibration check with Cytek QbSure beads.

PM performed by Cytek on 10/24/2018 - complete instrument operation verified.

The testing performed verifies operation by requiring the instrument to pass calibration. In order to have a passing result, the fluidics system has to work perfectly to have the correct flow rate. The sampling system has to deliver the calibration beads to the flowcell for calibration to be started. The lasers and optics must be clean, aligned with working detectors and properly operating fluidics to get a passing result.

System consists of:
BD FACSCalibur Flow Cytometer
HTS Microplate Sampler Module
FACSFlow Supply
MiniMac Desktop PC with CellQuest Pro software ver6.1 w/ License dongle and PlateManager software ver2.1 (with monitor, keyboard, mouse)
Communication and power cables.

Configuration and Setup:
Excitation lasers: Blue 488 nm and Red 635 nm
Sample types: Cells and multiplex bead assays
Optics: FSC, SSC, FL1(530/30nm), FL2(585/42nm), FL3(670nm LP) and FL4(661/16nm)
Applications: Cell counting, aiding DNA analysis, Bead Array assays and Immunophenotyping

Video Description:
We zoom out to get a view of the front and sides of the BD FACSCalibur Flow Cytometer.
From here we move across the computer and the HTS to the monitor to do a software and status check.
We then move back to the BD FACSCalibur Flow Cytometer to get a look at the internals of the system.
Following this we go around to the back side of the BD FACSCalibur Flow Cytometer getting a clear look at the serial tags.
Returning to the front, we get a wide view of the BD FACSCalibur Flow Cytometer.

Documentation:
Datasheet: https://drive.google.com/file/d/0Bz5srQzgsa6uRmtfdTZKSEJDRlE/view?usp=sharing

Brochure: https://drive.google.com/file/d/0Bz5srQzgsa6udjZralZ4UGFZdEU/view?usp=sharing

Manual: https://drive.google.com/file/d/0Bz5srQzgsa6uSUNiN1ZDNnNuNEE/view?usp=sharing

This will ship via freight. Please contact us with your zip code to obtain shipping quote.

Please do not hesitate to contact us if you have any questions.

 photo 14075- 1.jpg photo 14075- 2.jpg photo 14075- 3.jpg photo 14075- 4.jpg photo Cytek bead test.png photo 14075- 5.jpg photo 14075- 8.jpg photo 14075- 9.jpg photo 14075- 10.jpg photo 14075- 11.jpg photo 14075- 12.jpg photo 14075- 13.jpg photo 14075- 14.jpg photo 14075- 15.jpg photo 14075- 16.jpg photo 14075- 17.jpg photo 14075- 18.jpg photo 14075- 19.jpg photo 14075- 20.jpg photo 14075- 21.jpg photo 14075- 22.jpg photo 14075- 23.jpg photo 14075- 24.jpg photo 14075- 25.jpg photo 14075- 26.jpg photo 14075- 27.jpg photo 14075- 28.jpg photo 14075- 30.jpg photo 14075- 31.jpg photo 14075- 32.jpg photo 14075- 33.jpg photo 14075- 34.jpg photo 14075- 35.jpg photo 14075- 36.jpg photo 14075- 37.jpg photo 14075- 38.jpg photo 14075- 39.jpg photo 14075- 40.jpg photo 14075- 41.jpg photo 14075- 42.jpg photo 14075- 43.jpg photo 14075- 44.jpg photo 14075- 45.jpg photo 14075- 46.jpg photo 14075- 47.jpg photo 14075- 48.jpg photo 14075- 49.jpg photo 14075- 50.jpg photo 14075- 51.jpg photo 14075- 54.jpg

 


BD FACSCalibur Flow Cytometer w/ HTS Sampler -PM/Service done / Dongle Included.

Manufacturer Description and Specification:

Manufacturer Documentation:

Boston Industries, Inc. © 2019